1
|
Monopolar and bipolar electro-cautery Unit
|
400 W into 500 ohms cutting, 100 W into 500 ohms coagulation, blening in two steps, 400KHz frequency, with 1000 KHz bipolar frequency.
Standard Accessories-
Colour coded double pedal foot switch with receptacle, silicone rubber patient plate with detachable connection cable and punched rubber straps, reusable active electrode handle with set of 4 electrodes and silicone rubber insulated cable, reusable insulated bipolar forceps with silicone rubber insulated cable, reusable handswitch surgical pencil with one flat electrode
|
One
|
2
|
Vacuum Suction-cum-irrigation apparatus
|
Max (-) 600 mm Hg Vacuum, 20 l/min free flow rate, 600 mm Hg irrigation pressure, 7 l/min free flow rate, 2 L capacity suction collection bottle having overflow protection, irrigation bottle suiting saline/glycene bottles of 500 to 3000 ml capacity, with inbuilt voltage stabilizer
With the following Standard accessories – Suction-Irrigation unit with integral controls, fluid collection bottle with rubber cap and overflow cut off (1 set), Metal tube connections for irrigation bottle – air pressurization needle (1 no) and outflow cannula (1 no), Irrigation bottle holder (1 set), Silicon rubber connection tube (1 pair)
|
One
|
3
|
Medical Image Capture and Reporting System
|
Hardware - Control computer CPU with Core 2 duo Processor of 2 GB memory & 250 GB HDD, Monitor, mouse and keyboard, External Video/image capture card, 3 pedal foot-switch
Software – Easy to use, functionally rich, video capture and reporting software for laparoscopy unit. Image, video and audio data storage facility in digital format and retrieval facility
|
One
|
|