Laminar flow hood
(Horizontal)
|
Filter : HEPA/ ULPA - Filter assembly: According to ISO Class 5 ( ISO 14644-1:1999 ( E ) or EN1822 or equivalent requirements - Particle Retention: 0.3 μ & above - Efficiency: 99.99% - Real time air flow monitoring system should display the performance of the system - Internal Depth 2ft - Internal Height 2ft - Internal width: 4 ft No. of HEPA Filter 1 - Illumination : Built-in warm white, electronically ballasted, 5000k lighting, ero-flicker and instant start - Detail of UV Tube : 1 x30 W - Outer Body : Complete Body made of M.S. sheet duly powder coated - Working Table : Made of single sheet S.S. Sheet satin finished (SS 304) - Media : Glass Fiber paper - Pre Filter: Media : Non woven synthetic Polyester -Particle Retention : 10-15 μ - efficiency : 90% - Cleanliness Level : Class 100 - Velocity : 70 FPM 20 % - Noise Level : 65 db A 5% - Front Door : Made of clean perplex sheet - Power Supply : 220V, 50 Hz
|
1
|
2
|
Hot air oven
|
Digital display temperature controller (bottom heated) , Temp. range 50 to 300 ⁰C, Accuracy ±1⁰C –Double walled with glass wool insulation -Inner chamber stainless steel- Perforated adjustable shelves -Dimensions: 45 x 45 x 45 cm Capacity 95 L
|
1
|
3
|
Water bath (Serological)
|
Dimensions : 33 L x 30 x 18 cm Capacity 15-18 Suitable for 4 racks). Temp : Ambient to 100 ⁰C - Digital temp. display - Accuracy ±1⁰C - Double walled with SS and glass wool insulation- 220V
|
1
|
4
|
Incubator
|
Digital display of temp. Bottom heated- Temp range 30 to 110 ⁰ Dimensions: 45 x 45 x 45 cm Capacity 95 L Inner chamber SS –two shelves - 220V
|
1
|
5
|
Liquid nitrogen container
-35 litres
|
LN2 capacity 30-35 L Storage container Neck diameter 50 mm Height 650-700 mm Static evaporation loss rate less than 150 ml/day. No of canisters 6-8 (double layered) - Straw capacity (0.25 ml) single level -1500
|
2
|
6
|
Liquid nitrogen container
-42 litres
|
LN2 capacity 42 L Storage container Neck diameter 60-70 mm Height 650-700 mm Static evaporation loss rate less than 250 ml/day. No of canisters 6-8 (double layered) - Straw capacity (0.25 ml) double level -3000
|
1
|
7
|
Liquid nitrogen container
-7 litres
|
LN2 capacity 7-8 L Storage container Neck diameter 50 mm Height 500-600 mm Static evaporation loss rate less than 150 ml/day. No of canisters 6-8 (double layered) - Straw capacity (0.25 ml) single level -1500
|
1
|
6
|
Liquid nitrogen container
- 11 litres
|
LN2 capacity 11-12 L- Storage container - Neck diameter 50 mm - Height 600-700 mm Static evaporation loss rate less than 150 ml/day. No of canisters 6-8 (double layered) Straw capacity (0.25 ml) single level -1500
|
6
|
7
|
Liquid nitrogen container
1 litre
|
LN2 capacity 1L- Storage container. Neck diameter 30 mm - Height 400 mm Static evaporation loss rate less than 100 ml/day. No of canisters -1 Straw capacity(0.25 ml) -100
|
1
|
8
|
Autoclave
|
Vertical : Size 2’ height 1.5 ‘ width - Temp : 121 ⁰C – 138 ⁰C - Pressure 1.1 to 2.2 Kg/cm2 of steam pressure - Single door high pressure steam sterilizer with double / triple walled, steam jacket - Material of construction - Sterilizer chamber SS 316 - Door SS 316 - Jacket MS - Loading carriage SS 316 - Door Gasket : Silicon or better - Insulation: fiber glass resin bonded wool or better - Insulation cover: SS sheets - Spring loaded safety valves and automatic vacuum breaker for jacket - Removable plug screen for chamber drain - Safety lock for door: pressure lock safety device - Microprocessor based controller with time and temperature programmable by user, - LCD display to display process and set value at a glance - All safety measures like double locking safety device, automatic safety shut-off to prevent overheating of the chamber and safety pressure-release valves for the chamber etc. - Voltage :220 – 240 V AC 50 Hz input power supply
|
1
|
9
|
Analytical Balance
|
Capacity 220 g – Sensitivity 1 mg – For weighing lab chemicals – Wind shield / glass case should be included – Sensitivity drift (10 – 30 C) –Operating temp 5 to 40 C - 85 % Relative humidity or less –Vacuum Fluorescent Display - Power 220 V 50 Hz
|
1
|
10
|
Magnetic stirrer cum hotplate
|
Magnetic Stirrer with inbuilt hot plate - Temperature: 0 - 250 °C with accuracy of ± 5 °C-Speed: 100 – 1100 RPM or more - Technology to ensure uniformity of the temperature - Working table should withstand high temperature, corrosion and chemical resistant - Input Voltage 230V/50Hz.A.C. Accessories: Magnetic stirring bars (cylindrical 5 to 10 cms), Holding rods, supporting rods.
|
1
|
11
|
pH meter
|
Bench type –pH slope and offset display-Range 0 to 14 Accuracy - ±0.01 pH –No. of calibrations –at 3 points –Temp compensation –automatic (0 to 100 ⁰C)- Temp Range 0 to 100 ⁰C – Temp accuracy - ±0.3⁰C- Electrode stand – integrated with swivel arm -Power supply AC/DC 9 V Adapter (220 VAC, 500 mA)
|
1
|
12
|
Micropipettes
|
Variable Volume ranges -1-10 µl, 10-100 µl, 20-200 µl, 100-1000 µl. Fully autoclavable pipettes- Easy to read display – Light and steady pipetting action.
|
1 each
|
13
|
Submarine gel electrophoresis system with accessories
|
Size: 125 x 150 mm -With removable and UV transparent gel casting tray - Complete system should be light weight, convenient with leak proof gel casting tray, Bubble Level, adjustable height- Combs to fit variable needs (1 mm, 1.5 and 2.0 mm thick well comb) – Buffer tank must be moulded - Size: 125 x 150 mm. The electrodes (terminals) should not be projected up right (It should be pit type in to which the cord can be fit into - Gel Caster - universal gel caster - Gel casting system for casting multiple gels -Platinum Electrodes.
|
1
|
14
|
Electrophoresis power packs
|
: Volts 0-300 V -Fully adjustable in 1V steps- Current 0-400 mA fully adjustable in 1 mA - Power 75 W or more - Type of out put:- Constant current, Voltage - Timer 1-99 hrs - Programmable Power settings - Automatic recovery after power failure - Pause/ resume run function - LED Display
|
1
|